When the temperature goes over 98 degrees while
you are in any of the procedure screens
When the auto stop button is pressed while in
motor or sensory stimulation with the stimulation
output switched on
When the auto stop button is pressed while in any
lesion or pulse mode with the RF output switched
on manually
When the auto stop button is pressed while in
lesion or pulse mode with the RF output switched
on with the auto start function
When manually starting the RF output in any
mode with the impedance of probe 1 open circuit
When manually starting the RF output in any
mode with the impedance of probe 2 open circuit
When manually starting the RF output in any
mode with the impedance of probe 3 open circuit
When automatically starting the RF output in
lesion mode with the impedance of probe 1 open
circuit.
Document 109.00
NEUROTHERM RADIO FREQUENCY LESION GENERATOR
MODEL NT1100
OPERATORS MANUAL
®
High temperature shutdown
Machine set to safe state
Emergency stop pressed
Motor stimulation cancelled and machine set to
safe state
Turn off Stimulate control to go back to procedure
screen
Emergency stop pressed
RF output cancelled and machine set to safe state
Turn off RF control to go back to procedure screen
Emergency stop pressed
RF output cancelled and machine set to safe state
Press lesion to go back to procedure screen
Unable to deliver RF power
Impedance open circuit on probe 1.
Turn off RF power control, check probe and
connection of the reference plate and restart
procedure.
Unable to deliver RF power
Impedance open circuit on probe 3.
Turn off RF power control, check probe and
connection of the reference plate and restart
procedure.
Unable to deliver RF power
Impedance open circuit on probe 3.
Turn off RF power control, check probe and
connection of the reference plate and restart
procedure.
Unable to deliver RF power
Impedance open circuit on probe 1.
Check probe and connection of the reference plate
and restart procedure.
Press lesion to go back to procedure screen