Custom Step Profiles - NeuroTherm NT1100 Operator's Manual

Radio frequency lesion generator
Table of Contents

Advertisement

Custom Step Profiles

The custom step profile facility gives the user the ability to create three lesion profiles.
Custom step profiles are only supported by single electrode or dual Electrode. The NT1100
defines a step profile using the following variables:
Start Temperature: The initial temperature in degrees centigrade that the profile should hold
Step Time:
Step Rise:
Final Temperature: The temperature in degrees centigrade that the profile should hold at after the
Final Time:
Start
Temp
From the graph, the key profile variables are identified as:
Start Temperature:
Step Time:
Step Rise:
Final Temperature:
Final Time:
Document 109.00
NEUROTHERM RADIO FREQUENCY LESION GENERATOR
MODEL NT1100
OPERATORS MANUAL
before the making the first step rise.
The time in minutes the profile should hold at each step temperature.
The increase in Temperature in degrees centigrade of each step.
last step rise.
The time in minutes that final temperature is held at before the end of the lesion
for this example we will create a custom step profile of the following:
Step time
50°C
2 Mins
5°C
65°C
4 Mins
®
Final time
Step rise
10September2012 Issue 06
7-20
Final temp.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents