NeuroTherm NT1100 Operator's Manual page 49

Radio frequency lesion generator
Table of Contents

Advertisement

C
HANGE SITE LABLELS DURING A SINGLE
In singe electrode mode, the site labels should be all updated before the procedures are
begun, then the Doctor can cycle through the labels during stimulation and lesions.
Note: While single electrode supports up to eight levels, the bluetooth printer only
recognises the first three site labels. Thus if the Doctor primarily used the printed reports
for documenting the treatment, it is recommended that only site labels 1-3 are used.
a) The Doctor can now update the site labels as described above for all sites to be
lesioned. Proceed with stimulation and lesion modes for the first site
b) Press the Up Select and Down Select buttons to cycle through each of the site labels.
They will appear on the screen as they are selected and stimulation is performed.
c) Select the End Session button
To return to the Procedure Settings screen
Alternative method to update site labels
e) Update site labels for all sites to be lesioned
f) Stimulate all sites sequentially by pressing Select Up and Select Down buttons to cycle
through each site. The site labels will appear on the screen as the Up/Down buttons are
selected.
g) Proceed to lesion mode, again using the up/down buttons to move from site to site
h) At the end of the session press the End Session button .To return to the Procedure
Settings screen.
Change site labels during other procedures
When performing bi-lateral procedures or ones that require additional sites with multiple
electrodes, these site labels can be updated during the patient's session without
adversely affecting the printed records or exported files
a) Update site labels as described for the first set of sites
b) Proceed with stimulation and lesion modes for those sites
c) Select End Session.
d) Select Site Labels, and update the labels to the new site locations
e) Select Accept and the procedure settings will re-appear
f) Select Accept and continue with stimulation and lesion for the second set of sites.
g) At the end of the session select End Session to return to Procedure Settings
Document 109.00
NEUROTHERM RADIO FREQUENCY LESION GENERATOR
MODEL NT1100
OPERATORS MANUAL
®
-
ELECTRODE PROCEDURE
10September2012 Issue 06
7-5