NeuroTherm NT1100 Operator's Manual page 34

Radio frequency lesion generator
Table of Contents

Advertisement

Setting Operating Parameters
The right hand side of the screen shows a series of five coloured boxes, each representing
the different operational states, Sensory Stimulation, Motor Stimulation, Thermal Lesion,
Pulsed RF and Pulsed Dose To enter each of the operational state menus, touch the relevant
coloured box using the touch screen. This will then prompt the user with a set of menus,
where key parameters are entered and stored under the Doctor profile.
Operating Parameters – [Defaults shown in bold]
Frequency [10,20,50,75,100,150,180,200]
Volts/Current [0-0.5V, 0-3.0V,5.0 V. 0-1.0mA,0-6.0mA, 0-10.0mA]
Thermal Lesion
Time[0:30,1:00,1:30,2:00,3:00,4:00,5:00,6:00,7:00,8:00,9:00,10.00]
Step Profile [None, IDET P1, IDET P2 IDET P3,custom 1,2 and 3
Set Temp [42,45,50,55,60,65] Time
[0:30,1:00,1:30,2:00,3:00,4:00,5:00,6:00,7:00,8:00,9:00,10:00,20:0]
Rate [1,2, 5,10]
Width [5,10,20,50]
Set Volts (30v,35v,40v,45v,50v,55v 60v,65v,70v, ,
Document 109.00
NEUROTHERM RADIO FREQUENCY LESION GENERATOR
OPERATORS MANUAL
Sensory Stimulation
Width [0.1,0.2,0.5,1.0]
Set Temp [50,55,60,65,70,75,80,85,90]
Pulsed RF
®
MODEL NT1100
Motor Stimulation
Frequency [2,5]
Width [0.1,0.2,0.5,1.0]
Volts/Current [0-0.5V, 0-3.0V,5.0 V]
Custom Step Lesion
Start Temp [50,55,60]
Step Time [0.01 0:10,0:30,1:00,2:00, 3:00]
Step Rise [1,5,10ºC]
Final Temp [65,70,75,80,85,90]
Final
Time[1:00,2:00,4:00,6:00.8:00,10:00]
Pulsed Dose
Set Temp [42]
Count [no set,120,240,360,480,600,1200]
Rate [2] Width [20]
Set Volt/Current
[30v,35v,40v,45v,50v,55v 60v,65v,70v,]
10September2012 Issue 06
6-7