IMPEDANCE
Measuring Frequency
Measuring Source Voltage
Measurement Display
Accuracy
Features
STIMULATION MODE
Signal Shape
Output Range
Pulse Rates
Pulse Rate Accuracy
Pulse Widths
Pulse width Accuracy
Document 109.00
NEUROTHERM RADIO FREQUENCY LESION GENERATOR
OPERATORS MANUAL
Voltage
Current
Motor
Sensory
®
MODEL NT1100
53 KHz (± 3KHz)
Less than 500 mV AC
50-2000 ohms (one ohm resolution)
± 5%
(a)
Internal 500 ohm Test Resistor
(b)
Impedance in all Lesion Modes and in Stimulation
Mode when stimulation is off
(c)
Audible Tone available where frequency varies
with impedance over full impedance range (50-
2000 ohms). Audible tone is adjustable and
mutable
(d)
Warning on screen if impedance is less than 50
ohms or greater than 2000 ohms
Biphasic square wave with negative edge leading. This
wave is available in a variety of frequencies and widths.
0-5v ± 3% for motor frequencies (2Hz and 5 Hz)
0-3v ± 5% (Default) for all other frequencies
0-0.5v ± 10% for all other frequencies
0-10mA ± 5% 50-2000 ohms
0-6mA ± 5% 50-2000 ohms
0-1mA ± 5% 50-2000 ohms
2 or 5 Hz (Default 2Hz)
10, 20, 50, 75, 100, 150, 180, 200 Hz (Default 50 Hz)
± 3%
0.1, 0.2, 0.5 and 1.0 mS (Default 1.0 mS)
± 5% for 0.2, 0.5 and 1.0 mS
± 15 % for 0.1 mS
10September2012 Issue 06
3-2