The Simplicity lll procedure is complete after the tone sounds and the RF is switched off
See Addendum on the NT1100 Section Number 10-11 for additional detail on the Simplicity ll and
Simplicity lll Functionality.
7.10
Three electrodes RF
1. As with Two electrodes RF, Three electrodes RF are only available in Auto mode. It is selected at
the "Procedure Setting" menu. Highlight the choice 'Three electrodes ' and press 'Accept'
2. Proceed from the procedure selection screen by pressing 'Accept'. The first step is the motor
and sensory stimulation. Each electrode has its own graphical area. Electrode 1 on the left, with
electrode 2 in the centre, with electrode 3 on the right. As with the two electrodes RF mode,
navigate between the electrodes using the