NEUROTHERM RADIO FREQUENCY LESION GENERATOR

Back Panel Layout

7
6
1.
Mains On/Off Switch
Is a rocker type switch, combined with an I.E.C. connector socket with twin 'in-line' anti-
surge fuses in a single unit to IEC 950
2.
Mains IEC Connector
The three pin plug of the mains must be pushed into this socket. This cannot be done
incorrectly i.e. with the live and neutral reversed because of the orientation of the unused
earth pin.
3.
Fuses and Voltage Changes
The NeuroTherm is protected by two in-line fuses, one on the live line and one on the
neutral line. These fuses are located to the right hand side of the IEC socket. The fuses
are 20mm Anti-Surge to BS 4265. 1 amp for 230v supply, 2 amps for 115v supply. To
access the fuse holder lift protective lid from the right hand edge and hinge back, the fuse
carrier can then be removed. The mains input unit also contains a small printed circuit
card which allows the mains input voltage to be changed [Note this is for factory setting
only and should not be altered]
4.
Serial Plate
This plate gives information on Rated Supply, Rated Power, Fuse Ratings and the
Machine Serial Number.
Document 109.00
®
MODEL NT1100
OPERATORS MANUAL
A
5
B
C
D
2
1
10September2012 Issue 06
5-7
4
3