NeuroTherm NT1100 Operator's Manual page 72

Radio frequency lesion generator
Table of Contents

Advertisement

Level 3 - User warning
Message Scenario
When trying to print, but there is an error with the
printer.
E.g. out of paper / out of ink / paper jam
When trying to switch into sensory/motor stim with
the stimulation pot switched on
When trying to switch into lesion/pulse mode with
the RF power control switched on
When the accept button is pressed in the setup
confirmation screen with No thermocouple elect
selected
Open circuit condition on Probe 1 with stimulate
control turned on
Open circuit condition on Probe 2 with stimulate
control turned on
Open circuit condition on Probe 3 with stimulate
control turned on
Impedance lower than 50 Ohm condition on Probe
1 with stimulate control turned on
Impedance lower than 50 Ohm condition on Probe
2 with stimulate control turned on
Impedance lower than 50 Ohm condition on Probe
3 with stimulate control turned on
Document 109.00
NEUROTHERM RADIO FREQUENCY LESION GENERATOR
MODEL NT1100
OPERATORS MANUAL
®
Message Text
Printing error - check printer
Fix problem, press <retry> or press <cancel> to
abort the print.
Turn off stimulation control before continuing
Turn off RF power control before continuing
No thermocouple elect selected.
Be aware that there will be no temperature
monitoring during the procedure.
Unable to stimulate probe 1
Impedance probe 1 open circuit.
Turn off stimulate control and check probe and
reference plate.
Unable to stimulate probe 2
Impedance probe 2 open circuit.
Turn off stimulate control and check probe and
reference plate.
Unable to stimulate probe 3
Impedance probe 3 open circuit.
Turn off stimulate control and check probe and
reference plate.
Unable to stimulate probe 1
Impedance probe 1 too low (<50 ohms)
Turn off stimulate control, check probe placement
and leads.
Unable to stimulate probe 2
Impedance probe 2 too low (<50 ohms)
Turn off stimulate control, check probe placement
and leads.
Unable to stimulate probe 3
Impedance probe 3 too low (<50 ohms)
Turn off stimulate control, check probe placement
and leads.
10September2012 Issue 06
8-5
Message
code
3-001
3-003
3-004
3-005
3-006
3-007
3-008
3-009
3-010
3-011