Rf Pulse Mode - NeuroTherm NT1100 Operator's Manual

Radio frequency lesion generator
Table of Contents

Advertisement

RF Pulse Mode

From the welcome screen select the "Accept" button to show the "Procedure Settings" screen
1. To access the Pulsed RF set up Menu, press the blue 'Pulse RF' box on the right hand side
of the display. Set Temperature, time, Pulse Rate, pulse width, and voltage required
2.
Once all parameters are set press the 'Accept' Button. To progress to an active state.
Carry out the Stimulation and Motor sensory checks and log the results by selecting the
"Log" Button
To switch the unit to RF Pulse mode, press the
appear.
For Pulsed RF in manual mode, turn the RF Control clockwise to increase the output power. For
Automatic mode the RF Control needs to be fully anticlockwise and then press the 'Auto Start'
button.
Once 2.00 minutes has passed at the set temperature, the alarm will sound. The procedure is
now complete.
Note In pulsed RF Mode a thermal lesion is not produced and the temperature is does not have
to reach the set temperature (See notes on Pulsed RF)
Document 109.00
NEUROTHERM RADIO FREQUENCY LESION GENERATOR
MODEL NT1100
OPERATORS MANUAL
®
button. The Pulsed RF screen will
10September2012 Issue 06
7-6