Earth Leakage - NeuroTherm NT1100 Operator's Manual

Radio frequency lesion generator
Table of Contents

Advertisement

Earth Leakage

1
Enclosure leakage current
Normal
Reverse
Single Fault Condition
Normal
Reverse
2
Patient leakage current
Normal (AC)
Reverse (AC)
Single fault condition
Normal (AC)
Reverse (AC)
3
Patient Leakage current
Normal (DC)
Reverse (DC)
Single fault condition
Normal (DC)
Reverse (DC)
4
Patient Auxiliary Leakage Current
Normal (AC)
Reverse (AC)
Single Fault Condition
Normal (AC)
Reverse (AC)
5
Patient Auxiliary Leakage Current
Normal (DC)
Reverse (DC)
Single Fault Condition
Normal (DC)
Reverse (DC)
6
Patient Leakage Floating Type
Normal
Reverse
Single Fault Condition
Normal
Reverse
Document 109.00
NEUROTHERM RADIO FREQUENCY LESION GENERATOR
OPERATORS MANUAL
®
MODEL NT1100
40 microamps
40 microamps
40 microamps
40 microamps
5 microamps
4 microamps
7 microamps
7 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
6 microamps
6 microamps
4 microamps
4 microamps
4 microamps
4 microamps
27 microamps
27 microamps
36 microamps
35 microamps
Typical
Maximum Allowable
100 microamps
100 microamps
500 microamps
500 microamps
100 microamps
100 microamps
500 microamps
500 microamps
10 microamps
10 microamps
50 microamps
50 microamps
100 microamps
100 microamps
500 microamps
500 microamps
10 microamps
10 microamps
50 microamps
50 microamps
5000 microamps
5000 microamps
5000 microamps
5000 microamps