UV-Radiation (UV-lamps only)
Emissions of ultraviolet radiation (200-315 nm) from this product is limited
such that radiant exposure incident upon the unprotected skin or eye of
operator or service personnel is limited to the following TLVs (Threshold Limit
Values) according to the American Conference of Governmental Industrial
Hygienists:
Table 38
Exposure/day
8 hours
10 minutes
Typically the radiation values are much smaller than these limits:
Table 39
Position
Lamp installed, 50-cm distance
Lamp installed, 50-cm distance
Agilent 1260 FLD User Manual
UV-Radiation Limits
Effective Irradiance
0.1 µW/cm2
5.0 µW/cm2
UV-Radiation Typical Values
UV-Radiation (UV-lamps only)
Effective Irradiance
average 0.016 µW/cm2
maximum 0.14 µW/cm2
12
Appendix
225