System Wire Sizes & Kva Load Characteristics; Jedi Generator 3-Phase 50 Kw System - Minimum Wire Size - GE Optima XR642 Pre-Installation

Table of Contents

Advertisement

GE Healthcare
Direction 5743854-1EN, Revision 1
Table 5-1 JEDI Generator Power Specifications
HV cable type
Ground Wire
2.2
System Wire Sizes & kVA Load Characteristics
• Calculations based upon nominal voltage, wire size in AWG. To convert to mm2, refer to Table 5-2.
• Recommended feeder sizes from distribution transformer to the power cabinet.
• Neutral must be terminated inside the main disconnect panel and not at any GE cabinet.
• The grounding conductor will be of same size as the feeder wires. This ground will run from equipment back to the
facility power source / main grounding point and always travel in the same conduit with the feeders and neutral.
2.2.1

JEDI Generator 3-Phase 50 kW System - Minimum Wire Size

Table 5-2 AWG Wire Size Conversion to mm
American Wire Gauge (AWG)
Table 5-3 JEDI Generator 3-Phase 50 kW System - Minimum Wire Size
WIRE RUN
LENGTH
15m (50 ft.)
30m (100 ft.)
46m (150 ft.)
61m (200 ft.)
77 m (250 ft)
92 m (300 ft)
107m ( 350ft)
122m (400 ft)
Page 98
USA: 22mm DSI (<= 165 pF/m)
HV cable connector = Federal standard
Same as power cable
2
Diameter (Inches)
6
0.1620
5
0.1819
4
0.2043
3
0.2294
2
0.2576
1
0.2893
1/0
0.3249
2/0
0.3648
3/0
0.4096
4/0
0.46
250M
0.575
300M
0.630
350M
0.681
400M
0.728
380
400
+/- 10%
+/- 10%
*6
*6
5
5
4
4
3
4
2
2
1/0
1
2/0
2/0
3/0
2/0
Section 2.0 - Electrical Requirements
Optima XR642/XR648 Pre-Installation
Diameter (mm)
4.11
4.62
5.19
5.83
6.54
7.35
8.25
9.27
10.40
11.68
14.6
16.0
17.3
18.49
INPUT VOLTAGE (VAC)
420
440
+/- 10%
+/- 10%
*6
*6
*6
*6
4
4
4
4
2
3
1
2
1/0
1
2/0
1/0
mm
Cross Sectional Area (
13.30
16.77
21.15
26.65
33.61
42.39
53.46
67.40
84.97
107.16
126.68
152.0
177.35
202.68
460
480
+/- 10%
+/- 10%
*6
*6
*6
*6
4
4
4
4
3
4
2
2
1
1
1/0
1/0
2
)