Error Messages

"WARNING: The EEPROM data was found to be either INVALID or
uninitialized. GE factory defaults will be stored in both the EEPROM
and the monitor's configuration memory. You will be required to re-
enter the network configuration, re-enable any password protected
features and restore all monitor settings and site-specific defaults."
"WARNING: THIS VERSION OF BOOT CODE IS NOT
COMPATIBLE WITH THE VERSION OF MAIN CODE CURRENTLY
STORED IN FLASH. PLEASE UPDATE THE BOOT CODE."
"Boot Flash test FAILED."
"ERROR: The Boot Code stored in Flash is not valid. Main Code
cannot be loaded until valid Boot Code exists."
"Main Flash test FAILED."
"ERROR: The Main Code stored in flash is not valid."
"Static RAM test FAILED."
"ERROR: The SRAM memory test failed. Main Code will not be
loaded unless this test passes. Reboot the monitor to repeat testing."
"Real Time Clock FAILED - will not start."
"WARNING: The real time clock chip is not running. Main Code
cannot be loaded until this chip is started. Attempting to start real time
clock..."
Followed by either:
"The real time clock was started. Select Start Patient Monitoring to
load and execute Main Code."
Or
"ERROR: Unable to start the real time clock."
PREVENTIVE MAINTENANCE REQUIRED
Internal lithium battery is LOW.
This message displays during power up. After the power-up tests, the
following message displays in the BOOT LOADER program:
ONE OR MORE POWER-UP TESTS HAVE FAILED ** SERVICE
MAY BE REQUIRED **
DO YOU WISH TO CONTINUE MONITORING ANYWAY?
6-6
Troubleshooting: Problems
The following table describes error messages that may appear and how to
resolve the problem.
Message
Transport Pro
Possible Reason/Solution
1. Restore Ethernet address and IP address.
2. Power cycle.
3. Access Boot Loader and restore all monitor settings and site-
specific defaults via Set Configuration and Options Menu.
4. If error message persists, replace the main processor FRU.
1. Power cycle.
2. Reload Boot Code.
3. If problems persists, replace the main processor FRU.
1. Power cycle.
2. If problems persists, replace the main processor FRU.
Perform all of the maintenance procedures.
The main processor battery may be depleted. Remove this
monitor from service and contact GE.
The battery switch on the main processor FRU may be off.
If problems persists, replace the main processor FRU.
2012659-042