Totalizer; A.3.4 Totalizer - Siemens SITRANS FC410 Operating Instructions Manual

Coriolis flowmeters with modbus
Hide thumbs Also See for SITRANS FC410:
Table of Contents

Advertisement

Modbus holding registers
A.3 Modbus holding registers tables
Modbus
Data type /
address
Size [bytes]
2442
Float / 4
2444
Float / 4
A.3.4

Totalizer

Table A- 7
Totalizer
Modbus
Data type /
address
Size [bytes]
2609
unsigned /
2
2610
float / 4
126
Parameter
Description
Density Correction
Set density compensation
Factor
value (gain) in order to
make a density correction
(scale factor).
To increase the displayed
density value with +0.5 %,
set the density factor to
1.005.
The displayed density value
will now be 0.5 % higher
than before
Density Correction
Set density compensation
Offset
value (offset) in order to
make an offset on the mea-
sured density.
To make the flowmeter
show + 2 kg/m
density offset to 2.000 kg/m
in the 'Sensor' menu
Parameter
Description
Totalizer State
Totalizer state
0 = paused
1= running
Totalizer Value
The totalized MASS value in
kg
Default value
[unit]
1
0 [kg/m
]
3
, change the
3
3
Default
value [unit]
1
0 [kg]
Operating Instructions, 02/2016, A5E33120874-AC
Value range Access
level
-1.999 to
Read /
+1.999
Write
-1 400 to +1
Read /
400
Write
Value range
Access
level
0 to 1
Read only
Min. -1.70E+38
Read Only
Max. 1.70E+38
FC410 with Modbus