Alarms; A.3.9 Alarms - Siemens SITRANS FC410 Operating Instructions Manual

Coriolis flowmeters with modbus
Hide thumbs Also See for SITRANS FC410:
Table of Contents

Advertisement

Modbus holding registers
A.3 Modbus holding registers tables
A.3.9

Alarms

Table A- 19
Alarms
Modbus
Data type /
address
Size [bytes]
3012
Unsigned /
4
136
Parameter
Description
Alarm Group 1
The following bit is set in case of
active alarm:
Bit 4:
Power Supply Malfunction
Bit 6:
Temperature Circuit
Malfunction
Bit 10:
Measurement Out Of Range
Bit 14:
Calibration Malfunction
Bit 15:
Compensation Out Of Range
Bit 17:
Pickup Malfunction
Bit 23:
Driver Malfunction
Bit 26:
Measurement Out Of Range
Bit 27:
Mass Flow Max Limit
Exceeded
Bit 28:
Volume Flow Max Limit
Exceeded
Bit 29:
Density Max Limit Exceeded
Bit 30:
Min Tube Temp Exceeded
Bit 31:
Max Tube Temp Exceeded
Default value
Value range Access
[unit]
-
-
Operating Instructions, 02/2016, A5E33120874-AC
level
Read Only
FC410 with Modbus