8.1.6

Totalizer

Totalizer function
The device has one totalizer function that can be used to totalize the mass flow process value.
Default unit of the totalizer can be changed, see Modbus register 8320.
The totalizer may be paused, resumed or reset:
● Pause (Modbus address 2613): the totalizer holds the last value before the failure occurred
● Resume (Modbus address 2614): the totalizer continues counting the actual measured
value
● Reset (Modbus address 2612): the totalizer continues counting based on the last input value
(for example mass flow) before the failure occurred.
Note
The totalizer is reset on power failure.
8.1.7

Access management

All parameters may be viewed, but some are protected against changes by access level
control.
The Access Management menu makes it possible to gain access to PIN code protected
parameters and to change PIN codes.
Figure 8-1
SITRANS FC310
Operating Instructions, 01/2019, A5E44036384-AB
Access Management menu
Parameter assignment
8.1 Functions
81