Siemens SITRANS FC410 Operating Instructions Manual page 127

Coriolis flowmeters with modbus
Hide thumbs Also See for SITRANS FC410:
Table of Contents

Advertisement

Table A- 4
Massflow
Modbus
Data type /
address
Size [bytes]
2125
Float / 4
2426
Float / 4
: See Sensor dimension dependent default settings (Page 139)
1)
Table A- 5
Volume flow
Modbus
Data type /
address
Size [bytes]
2170
Float / 4
: See Sensor dimension dependent default settings (Page 139)
1)
Table A- 6
Density
Modbus
Data type /
address
Size [bytes]
2127
Float / 4
2129
Unsigned /
2
FC410 with Modbus
Operating Instructions, 02/2016, A5E33120874-AC
Parameter
Description
Low Massflow Cut-
Set massflow limit for low
Off
flow cut-off.
Below this limit massflow
output is forced to zero.
If Low Flow Cut-Off is set to
0, the cut-off functionality is
disabled.
Notice:
It is recommended to set a
lower value for gas applica-
tions.
Massflow Correction
Specify correction factor for
Factor
use in the massflow calcula-
tion
Parameter
Description
Low Volumeflow
Define the numerical volume-
Cut Off
flow value below which the
volume flow output is forced
to zero.
Parameter
Description
Empty Tube Limit
Define threshold value of
empty tube
Empty Tube
Set automatic detection of
Detection
Empty Tube On/Off
0 = off (Empty tube is off).
1 = on (a density value
below Empty Tube Limit
triggers an alarm. All flow
rate values are forced to
zero %).
Modbus holding registers
A.3 Modbus holding registers tables
Default value
Value range Access
[unit]
Sensor size
0 to 1023
specific [kg/s]
1)
1
-1.999 to
+1.999
Default value
Value range Access
[unit]
Sensor size
0 to 0.177
specific [m
/s]
3
1)
Default value
Value range Access
[unit]
500 [kg/m
]
-14 000 to
3
+14 000
0
0 to 1
level
Read /
Write
Read /
Write
level
Read /
Write
level
Read /
Write
Read /
Write
125