Siemens SITRANS FC410 Operating Instructions Manual page 133

Coriolis flowmeters with modbus
Hide thumbs Also See for SITRANS FC410:
Table of Contents

Advertisement

Modbus
Data type /
address
Size [bytes]
2138
Float / 4
2140
Float / 4
2142
Float / 4
2144
Unsigned /
2
2145
Unsigned /
2
2180
Unsigned /
2
: See Sensor dimension dependent default settings (Page 139)
1)
FC410 with Modbus
Operating Instructions, 02/2016, A5E33120874-AC
Parameter
Description
Standard Deviation
Set limit for zero point ad-
Limit
justment Standard Deviation
value. If the Standard De-
viation exceeds the
Standard Deviation Limit,
the auto zero point adjust-
ment is aborted.
Zero Point Offsett
Set limit for zero point off-
Limit
set.
If the zero point offset
exceeds the zero point
offset limit, the zero point
offset cannot be stored
Zero Point Offset
Default zero point offset
Value
based on factory calibration
of sensor.
A Zero point offset compen-
sates for sensor variations
due to process conditions.
Zero Point Adjust
Shows the progress of the
Progress
currently running Zero Point
adjustment in percentage
Zero Point Adjust
Status of the last zero point
Status
adjustment performed
Every high bit ('1')
represents an error
occurred in the last zero
point adjustment performed.
No high bits equals ok.
Bit 0 = Zero sigma limit
exceeded
Bit 1 = Zero offset limit
exceeded
Bit 2 = Quality of zero point
conditions
Start Zero Point
Start automatic zero point
Adjustment
adjustment.
The automatic zero point
adjustment determines the
application specific zero
point offset automatically.
Possible selections:
0: Idle
1: Running
2: Start
Modbus holding registers
A.3 Modbus holding registers tables
Default value
Value range Access
[unit]
Sensor size
0 to +1023
specific [kg/s]
1)
Sensor size
-1023 to
specific [kg/s]
+1023
1)
0 [kg/s]
-1023 to
+1023
0 [%]
0 to 100
-
1
2
4
0
0 to 2
level
Read /
Write
Read /
Write
Read only
Read Only
Read Only
Read /
Write
131