Rotor Profile
a. Maximum speeds are based on a solution density of 1.2 g/mL with the following exceptions: the JA-18.1 rotor is rated for a
density of 1.4 g/mL; the JCF-Z rotor is rated for a density of 1.45 g/mL; and the JE-5.0 rotor is rated for a density of 3 g/mL.
b. Temperature performance for the JA-30.50 Ti rotor in the Avanti J-26XP is as follows: 15°C minimum at 26 000 rpm (at 30°C
ambient); 4°C minimum at 21 500 rpm (30°C ambient).
c. Maximum speed for the JA-30.50 Ti rotor is 26 000 rpm in the Avanti J-26XP centrifuge, and 30 000 rpm in the Avanti J-30I
centrifuge.
d. Temperature performance for the JA-25.50 and JA-25.15 rotors in the Avanti J-26XP is as follows: 10°C minimum at
25 000 rpm (at 30°C ambient); 4°C minimum at 21 500 rpm (30°C ambient).
e. Maximum speed for the JS-24.38 and JS-24.15 rotor is 10 000 rpm in the Avanti J-26XP centrifuge, and 24 000 rpm in the
Avanti J-30I centrifuge.
f. In the Avanti J-26XP, the JCF-Z Continuous Flow/Zonal Rotor can be used in the continuous flow and reorienting gradient zonal
configurations only.

Preinstallation Requirements

Do not attempt to install this instrument. Its purchase price includes installation by Beckman Coulter
personnel. Installation by anyone other than an authorized Beckman Coulter representative invalidates
any warranty covering the instrument.
Preinstallation requirements have been sent prior to shipment of the instrument. The following
information is provided in case the centrifuge must be relocated.
The centrifuge will be installed upon initial purchase by a Beckman Coulter Field Service
representative after preinstallation requirements for power and site preparation have been met.
The following equipment is required for preinstallation:
• Voltmeter
• For single phase centrifuges: two 30-ampere circuit breakers
• For three phase centrifuges: three 16-ampere circuit breakers
J326XP-IM-5AB
Rotor
Description
Code
JCF-Z
JCFZ
Continuous Flow/
f
Zonal Rotor
JE-5.0
5.0
Elutriation Rotor
r
max
large
chamber = 168 mm
standard
chamber = 125 mm
Sanderson
chamber = 126 mm
Specifications and Preinstallation Requirements
Preinstallation Requirements
Max
Max
RCF
a
RPM
( g)
Max Capacity
20 000
39 900
660 mL
(standard core)
1250 mL
(large core)
240 mL
(small core)
5 000
4 700
40 mL
3 500
4.0 mL
4 230
5.5 mL
1
Rotor
Manual
Number
JCFZ-IM
JE5-IM
1-7