Introduction

Preinstallation requirements have been provided for your Avanti J-E centrifuge. The following information
is included in case the instrument must be relocated. (After the centrifuge is moved it must be leveled by
adjusting the two front feet.)
NOTE
This centrifuge is designed to be installed by a Beckman Coulter Field Service representative.
Installation by anyone other than authorized Beckman Coulter personnel invalidates any warranty
covering the instrument.

Space Requirements

WARNING
Do not place the centrifuge near areas containing flammable reagents or
combustible fluids. Vapors from these materials could enter the centrifuge air
system and be ignited by the motor. Maintain a 30-cm (1-ft) clearance area around
the centrifuge while it is running. No persons or any hazardous materials
should be within this area while the centrifuge is operating except to change
operating controls, if required.
If it is necessary to move the centrifuge, maintain the following conditions.
1
Select a location away from heat-producing laboratory equipment, with sufficient ventilation
to allow for heat dissipation.
The centrifuge must have adequate air ventilation to ensure compliance to local requirements
for vapors produced during centrifuge operation.
2
Position the centrifuge on a level floor that can support the weight of the centrifuge and resist
vibration.
PN SJ-IM-8AB
Preinstallation Requirements
APPENDIX A
A-1