CONTROL PANEL
Quick buttons
5.2.3
Changing exposure values for quick buttons
16 Planmeca ProMax 3D Mid
The Select kV / mA display appears. Select the required
exposure values by touching the kV and mA values. The
selected values will be highlighted.
Confirm your selection and return to the main display by
touching OK.
The preset exposure values for the selected exposure
program are shown in the quick buttons at the bottom of
the display.
NOTE Selecting kV and mA values manually will override the
automatic quick button setting.
The quick button exposure values have been preset at the
factory. If needed, the preset values can be changed by
the user.
To change the preset values of a quick button, first select
the required kV and mA values by touching the
corresponding fields on the parameter display and then
touch and hold the desired quick button at the bottom of
the display until you hear a signal tone. The quick button
will now show the new exposure values.
The quick button values are program-specific, i.e. the
exposure values can be set separately for each program.
NOTE To view the preset exposure values for each program
type, refer to the tables in chapters 8 "PANORAMIC
EXPOSURE" on page 32, 9 "TEMPOROMANDIBULAR
JOINT (TMJ) EXPOSURE" on page 41 and 10 "SINUS
EXPOSURE" on page 53.
Confirm your selection and return to the main display by
touching OK.