5.2

General settings

5.2.1
Selecting patient size
From left to right: Child
5.2.2
Selecting kilovolt and milliampere values
User's Manual (2D)
Adolescent
Small adult
Average-sized adult
Large adult
NOTE The exposure values (kV and mA) will automatically
NOTE Jaw shape and size selection is not available in sinus
NOTE The exposure values will automatically change to
Select the patient size by touching the corresponding size
symbol in the Patient size field on the main display.
Select the smallest patient size symbol for a child.
Selecting
the
smallest
automatically reduce the exposure area and patient dose.
The jaw shape & size setting will automatically change to
Normal & Narrow.
Select the second patient size symbol for an adolescent.
Select the third patient size symbol for a small adult.
Select the fourth patient size symbol for an average-sized
adult. Selecting any of the patient sizes in the middle will
automatically change the jaw shape and size setting to
Normal & Average.
Select the biggest patient size symbol for a large adult.
Selecting the biggest patient size will automatically
change the jaw shape and size setting to Normal & Wide.
The selected patient size will be highlighted.
change to correspond with the selected patient size
and exposure program.
programs.
correspond with the selected patient size and
exposure program.
The preset exposure values are average values and they
are only meant to guide the user.
To change the preset exposure values, touch the kV / mA
field on the main display.