HP 1652B Programming Reference Manual page 240

Logic
Hide thumbs Also See for HP 1652B:
Table of Contents

Advertisement

DURation
Command Syntax:
where:
G T
LT
c duration-value 7
Example:
Query Syntax:
Example:
HP 16528/1653B
Programming Reference
The DURation command allows you to specify the duration qualifier to be
used with the pattern recognizer term in generating the timing trigger.
The duration value can be specified in 10 11s increments within the
following ranges:
Greater than (GT) qualification - 30 11s to 10 ms
l
Less than (LT) qualification - 40 ns to 10 ms.
l
The DURation query returns the current pattern duration qualifier
specification.
:: = greater than
:: = less than
:: = real number
OUTPUT XXX; ":MACHINE1:TTRACE:OllRATION
10 DIM D$[lOO]
20 OUTPUT XXX; ":MACHINE1:TTRACE:DURATION?"
30 ENTER XXX;D$
4 0 P R I N T O$
50 EN0
DURation
command/query
G T , 40.OE-9"

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 1653b

Table of Contents