PPOWer

Query Syntax:
Example:
HP 16528/16538
Programming Reference
The PPOWer (preprocessor power) query returns the current status of
the HP 1652BE3B's high-current limit circuit. If it is functioning properly,
0 is returned. If the current draw is too high, 1 is returned until the
problem is corrected and the circuit automatically resets.
:PPOWer?
{o 1 I}
10 DIM Response$[lO]
20 OUTPUT XXX;":PPOWER?"
30 ENTER XXX;
40 PRINT Response%
50 END