RUNTil

Command Syntax:
where:
< v a l u e >
Examples:
HP 16528/1653B
Progmmming Reference
The RUNTil (run until) command defines stop criteria based on the time
between the X and 0 markers when the trace mode is in repetitive. When
OFF is selected, the analyzer will run until either the "STOP" touch screen
field is touched or the STOP command is sent. Run until the time
between X and 0 marker options are:
Less Than (LT) a specified time value
l
Greater Than (GT) a specified time value
l
In the range (INRange) between two time values
l
Out of the range (OUTRange) between hvo time values
l
End points for the INRange and OUTRange should be at least 10 ns apart
since this is the minimum time at which data is sampled.
This command affects the timing analyzer only, and has no relation to the
RUNTil commands in the SLISt and COMPare subsystems.
The RUNTil query returns the current stop criteria.
: : = { O F F 1 LT,cvalues 1 GT,<value> 1 lNFhge<value>,<value>
:: = real number
OUTPUT XXX;":MACHINEl:TWAVEFORM:RUNTIL
OUTPUT XXX;":MACHINE1:TWAVEFORM:RLlNTIL
RUNTil
command/query
GT, 800.OE-6"