Component List - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Yes
BC Switch goes to 897 or more, even if only once! Read before performing! Return to
page 30
, procedure number
THREE (3) possible bad components (see COMPONENT LIST below):

COMPONENT LIST:

1) Bad Actuator-Test Board
restricted balancing chamber valves #32, #34, #36 and #38
No BC Switch is NEVER 897 or more. A Flow Error is not presenting at this time! It may be
extremely intermittent. Return to
1
To LOCATE the Actuator-Test board refer to
2
To LOCATE these valves refer to
F- 1.0.5
but if you return here, because of no drain flow,
1
OR; 2) Bad actuator cable OR; 3) Multiple intermittent
page 52
, procedure number
Figure 4A
Figure 20
(page 92)
LEFT BLANK INTENTIONALLY
2008T Troubleshooting Guide
P/N 490292 Rev. A
2
.
F-
3.8.0.
(page 10);
Page 91