Isolate Temp; Post Temp Within 1.2 Pre Temp/Isolate Post Dialyzer Cond - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

CO- 10.0.10 ISOLATE TEMP
a) Call debug screen 0. Figure right, ENSURE
Cond
(mS) is within 0.4 of TCD.
b) ENSURE Pre Dialyzer Temp is between 35.5 and 38.5° C.
c) Is Post Dialyzer Temp within 1.2° C of Pre Dialyzer Temp?
Yes
Post Temp within 1.2 of Pre Temp! S
CO- 10.0.11
number
No Post Temp IS NOT within 1.2 of Pre Temp! Proceed to
CO-
10.0.20.
CO- 10.0.11 POST TEMP WITHIN 1.2 OF PRE TEMP / ISOLATE POST DIALYZER COND
Post Dialyzer Cond changes but overall is it remaining within 0.4 mS of Pre Dialyzer Cond?
Yes
Post Cond within 0.4 of Pre Cond! Proceed to
No
Post Cond is NOT within 0.4 of Pre Cond! Proceed to
CO-
10.0.30.
Pre Dialyzer
ee procedure
(page 417).
page 423
LEFT BLANK INTENTIONALLY
2008T Troubleshooting Guide
P/N 490292 Rev. A
page 418
, procedure number
, procedure number
page 421
, procedure number
CO-
10.0.40.
Page 417