Fresenius Medical Care 2008T Troubleshooting Manual page 737

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Figure (previous page)
D) Per the
segment between (blue) Pressure Sensor #110 and the bibag Connector.
E) Is Bic Press now between negative 50 and positive 30 (Yes or No)?
Yes
Bic Press between negative 50 and positive 30! a) Reattach Tube #2! b) Figure below,
check for restrictions between Valve #104 and the bibag Air Separation Chamber AND
between the bibag Air Separation Chamber and the Bicarb Pump!
No
Bic Press is NOT between negative 50 and positive 30! Leaving Tube #2 off, see procedure
number
V104SC- 1.0.3
pull Tube #2 off the rear side of the bibag Connector i.e. the
(page 734).
NO TEXT INTENIONALLY
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 733