Return Systems/Temp Comp 2; Isolate Temperature Internal Limits - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

T- 1.3.70 RETURN SYSTEMS / TEMP COMP?
IMPORTANT!
a)
Plug NTC #2's connector into the
distribution board position from the left, "CON-NTC"!
IMPORTANT!
b)
Avoiding VACANT position #4, RETURN
NTC #3's connector to the
from the left, "MON-NTC".
c) Enter Service Mode → Options → Hardware Options. Does the "Yes" box at
'Temp Comp' have a blue 'X' in it? (Yes or No)
Yes
Place the 'X' in the "No" box and press 'Enter'. The 'X' turns blue!
Post a note that 'Temp Comp' is off then see procedure number
T- 1.3.80
No
Temp Comp = No! See procedure number
T- 1.3.80 ISOLATE TEMPERATURE INTERNAL LIMITS
a) Select Calibrate Sensors → Temp Control.
b) Connect a temperature meter to the dialyzer connectors.
c) Sharply press 'Enter' TWICE to turn the [TEMP DAC] data box pale yellow / white.
d) Press [TEMP DAC], it turns bright yellow.
e) Set [TEMP DAC] to "130" and press 'Enter'. ENSURE the [TEMP DAC] is pale yellow / white. If gray
exit the calibration then return to part a!
IMPORTANT!
f)
Allow five (5) minutes BEFORE continuing to part g!
g) Press 'Enter' twice to activate the [Monitor Reference] data box. It turns bright yellow.
h) IGNORING the external meter adjust the [Monitor Reference] data box to 34.0° C!
i)
Press and release 'Enter' twice to save the calibration.
j)
See procedure number
nd
2
distribution board position
(page 260).
T- 1.3.90
(page 261).
2008T Troubleshooting Guide
P/N 490292 Rev. A
st
1
T- 1.3.80
(page 260).
Page 260