CLEAN- 2.0.5 CANNOT ACHIEVE 130 ML
Figure right, TWO (2) possible scenarios:
1)
IF (and ONLY if) NOT equipped with a bibag
Connector: See procedure number
(page 187).
2) IF equipped with a bibag Connector: See parts a
THROUGH e below:
a) Figure below, clamp the tubing attached to the
top of Valve #101.
b) Again attach the 60 ml syringe, with the plunger pushed all the way in, to the Deaeration Pump's
CLEAR tubing.
c) Pull the plunger to the end of the barrel then remove the syringe.
d) From the tubing, MEASURE flow for one (1) minute, looking for at least 130 ml!
e) Repeat parts b THROUGH e up to
minute?
Yes
More than 130 ml per minute! THREE (3) possible problems: 1) Loose bibag Interface
Board ribbon cable (refer to
bibag Interface Board.
No
After ten (10) attempts you DO NOT measure more than 130 ml per minute! Remove the
clamp from Valve #101 then see procedure number
CLEAN- 2.1.0
TEN TIMES
OR until if you measure at least 130 ml in one (1)
Figure
4C, page 11)) OR; 2) Bad Valve #101 OR; 3) Bad
2008T Troubleshooting Guide
P/N 490292 Rev. A
CLEAN- 2.1.0