FIL- 3.5.0 -250 MMHG HELD / ISOLATE VALVE #43 FLOW
Return the dialyzer lines to the shunt and close
a)
the door!
From the Home screen, set [Dialysate Flow] to 300
b)
ml/min and press 'Enter'!
c) Figure right, separate the Air Sensor Connector's brown and blue
wires. This eventually causes a Filling Program!
d) Obtain a 1000 ml (or larger) graduated cylinder!
e) Call debug screen 1.
f)
Figure right, if a 'Quick Connector' is used, at the end
of the 'to drain' tubing an adaptor is required!
g) Measure drain flow for two (2) minutes. TWO (2)
possible scenarios:
1)
IF (and ONLY if) 600 ml or more collected: See procedure number
2)
IF less than 600 ml collected: ENSURING [Dialysate Flow] is set to 300 ml/min, perform parts a
AND b below:
a)
Figure below, CAREFULLY trace the wire harness from distribution board positions "X9, P-
Dial" AND "X10, CFS" to ENSURE they terminate at the correct Pressure Transducer. If NOT,
this may be the problem!
b)
Proceed to
Allow FILACT = 1!
page 49
, procedure number
2008T Troubleshooting Guide
P/N 490292 Rev. A
F- 3.1.3
FIL- 3.6.0
(page 471).