Post Temperature Reference More Than 20 - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Yes
[Post-Temperature Reference] more than 20! Without pressing 'Confirm' yet, see
procedure number
No
[Post-Temperature Reference] is less than 20! See parts a and b below:
a) ENSURE the
NOT, repeat (ABOVE) procedure number
using the
b) If SURE the plug is placed correctly proceed to
CO-
11.0.00.
CO- 10.0.24 POST TEMPERATURE REFERENCE MORE THAN 20
'Sharply' press 'Enter' ONCE. Does an "Error" banner appear?
Yes
"Error" banner appears! Read before performing! Turn the machine off then
CAREFULLY repeat (ABOVE) procedure number
34° C
using the
reoccurs see procedure number
No
"Error banner does NOT appear. See parts a through e below:
a)
The screen says "2. Connect a 5.117 K ohm resistor...".
b) Place the 41° C (5.117 KΩ) plug, from the FOUR-RESISTOR SET, into distribution board
position, "x44, NTC-POST".
c) The screen's [Post-Temperature Reference] box will be more than 150 if the 41° C plug is
placed properly.
d) Press 'Enter'. The screen says "3. Calibration saved" and the screen's [Post-
Temperature] data box should = 41° +/- 0.1 °C. Press 'Enter' again to save the
calibration...
e) Return to Dialysis Program ("Select Program" → 'Dialysis' → 'Enter')!
f)
See procedure number
CO- 10.0.24
(page 419).
34° C
(6.808 KΩ) plug is placed properly at position "x44, NTC-POST". If
34° C
plug from another FOUR-RESISTOR SET!
FOUR-RESISTOR SET
plug from another
CO- 11.0.00
CO- 10.0.26
LEFT BLANK INTENTIONALLY
2008T Troubleshooting Guide
P/N 490292 Rev. A
CO- 10.0.22
(page 418) AND/ OR consider
page 432
CO- 10.0.22
page 432
(
).
(page 420).
, procedure number
(page 418). CONSIDER
. If the "Error" banner
Page 419