FIL- 2.0.4 NO AIR SEEN / ISOLATE FILLING PROGRAM
Call debug screen 1. WITHOUT LOOKING AWAY watch FILACT
(middle column) for up to two (2) minutes. If EVER = 1, even just once,
indicates a Filling Program! TWO (2) possible scenarios:
1)
IF (and ONLY if) FILACT EVER = 1: Proceed to
2) IF FILACT ALWAYS = 0: See parts a THROUGH d below:
Open the shunt door and LEAVE IT OPEN till instructed otherwise!
a)
Place a 'dummy venous chamber' in the Level Detector module.
b)
c) Press and release 'Reset' key then immediately press and hold it for three (3) seconds. Allow thirty
(30) seconds. If any alarm reoccurs attempt reset up to twice more BEFORE continuing to part d.
d) Can you RESET ALL alarms?
Yes
All alarms reset! See procedure number
No Address the alarm that cannot be reset!
FIL- 2.0.5 ALARMS RESET
Call debug screen 0. Per the Figure below, WITHOUT LOOKING AWAY, watch the UF Pump's 'dot' for
forty five (45) seconds. Does it EVER BLINK between white and blue?
Yes
The UF 'dot' blinks! The Actuator-Test Board may be bad!
No
The UF 'dot' does NOT blink! See procedure number
page 464,
FIL- 2.0.5
2008T Troubleshooting Guide
P/N 490292 Rev. A
procedure number
(page 462).
FIL- 2.0.6
(page 463).
FIL-
2.1.0.
Page 462