T- 7.6.0 METER TEMPERATURE IS STABLE AT 37°C / CHECK NTC #3 STABILITY
The screen's [Monitor Reference] data box should be gray AND at this point it does not matter if it = 37°
C! WITHOUT LOOKING AWAY, watch it one (1) minute noting its highest and lowest value. TWO (2)
possible scenarios 1) or 2) below:
1)
IF (and ONLY if) [Monitor Reference] changes more than +/- 0.2° C: ENSURING the external
meter is STABLE between 36.9 and 37.1° C, proceed to
2) IF [Monitor Reference] does NOT change more than +/- 0.2° C: See parts a AND b below:
a) 'Sharply' press 'Enter' to turn the [Monitor Reference] data box pale yellow / white.
b) Activate [Monitor Reference], it turns bright yellow. Without pressing 'Enter' or 'Confirm' until
instructed can you adjust [Monitor Reference] to match the external meter?
Yes
[Monitor Reference] should be bright yellow AND between 36.9 and 37.1° C. See
procedure number
No
With the machine off, swap in the listed components (see COMPONENT LIST below), one
at a time, with known good, and in between return to (ABOVE) procedure number
(page 323) to test each new component.
COMPONENT LIST: 1) NTC #3*. *To LOCATE NTC #3 refer to
2) Functional Board.
T- 7.6.2 TEMPERATURE CHECKS / SLOW FLOW CALIBRATION?
CAREFUL HERE! This procedure determines if a 'Slow Flow' temperature calibration is necessary:
a) 'Sharply' press 'Enter'! [Monitor Reference] MUST return to pale yellow / white!
b) 'Sharply' press 'Enter' again. TWO (2) possible scenarios:
1)
IF the screen says "9. Connect an external temperature meter...": A 'Slow Flow' calibration
is required! See procedure number
2) IF the screen says "Calibration saved. Press CONFIRM": If the procedures were done correctly
the [Monitor Reference] data box is GRAY. Does it continue to read between 36.9 and 37.1° C?
Yes
The GRAY data box = between 36.9 and 37.1° C. This completes the calibration
HOWEVER, if you started out troubleshooting a "TEMP OVER 95 DEGREES" alarm in
Heat Disinfect proceed to
Dialysis Program.
No The GRAY data box DOES NOT = between 36.9 and 37.1. Swap in the listed components
(see COMPONENT LIST below), one at a time, with known good then return to (ABOVE)
procedure number
COMPONENT LIST: 1) NTC #3*. *To LOCATE NTC #3 refer
2) Functional Board.
T- 7.6.2
(page 329).
T- 7.6.4
(page 330).
page 265
, procedure number
T- 7.2.1