-250 Mmhg Holding/Isolate Dialysate Pressure (Pdial); Check Pdial To 520 Mmhg; Check Pdial Zero - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

TMP- 2.2.8 -250 mmHg HOLDING / ISOLATE DIALYSATE PRESSURE (PDIAL)
a) ENSURING the external meter remains about -250 mmHg, call
debug screen 0.
b) Figure right, if (and ONLY if) Valve #25's 'dot' is BLUE continue
to part c. If NOT blue, return to (ABOVE) procedure number
page 499
TMP- 2.0.17
(
c) Figure right, look at screen 0's
possible scenarios:
1)
IF (and ONLY if) PDial AGREES with the external meter within +/- 30: See procedure number
TMP- 2.3.0
(page 502).
2) IF PDial DOES NOT agree with the external meter within +/- 30: Proceed to
procedure number
TMP- 2.3.0 CHECK PDIAL TO 520 mmHg
a) PULL on the syringe plunger to create negative pressure.
b) Clamp the 'to syringe' tubing segment.
c) Remove the syringe then reattach it.
d) PULL again. Can you make screen 0's PDial reach at least -520 mmHg?
Yes
-520 at PDial! See procedure number
No
Cannot achieve -520! Proceed to
TMP- 2.3.1 CHECK PDIAL ZERO
a) Remove the syringe and ALL clamps to open the four way to atmosphere!
b) ENSURING the transducer protector, at the meter, is NOT wet, does screen 0's PDial = 0 +/- 40?
Yes
PDial = 0 +/- 40! See procedure number
No PDial IS NOT between 0 and 40! Proceed to
).
PDial
data window. TWO (2)
TMP-
6.0.0.
page 513
2008T Troubleshooting Guide
P/N 490292 Rev. A
TMP- 2.3.1
(page 502).
, procedure number
TMP- 2.3.2
(page 503).
page 513
, procedure number
page 513
,
TMP-
6.0.0.
TMP-
6.0.0.
Page 502