CO- 1.0.8.1 [CONDUCTIVITY] MORE THAN 14.5 mS / ISOLATE PUMP TUBING CONNECTIONS
a) TWO (2) possible scenarios to ENSURE the Acid and Bicarb pump tubing is NOT reversed!
1)
IF equipped with a bibag Connector: Per the Figures below, trace the clear (INPUT) tubing
ACID PUMP
from the
2)
IF NOT equipped with a bibag Connector: Per the top Figure below, trace the tubing from the
red (ACID) connector to ENSURE it goes to the input of the
b) If ABSOLUTELY SURE the pumps are plumbed correctly, see procedure number
(page 360).
to ENSURE it comes from (blue) Pressure Sensor #106.
2008T Troubleshooting Guide
P/N 490292 Rev. A
ACID PUMP!
CO- 1.0.8.2
Page 359