Debug Screen 5 Olc Values Table - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

NOTE: Table B readings are affected by changing the [Dialysate Flow] rate, or turning it off, Filling
Programs, Pressure Tests and Temperature or Conductivity alarms (i.e. bypass). ENSURE the
machine is stable before responding to variations!
Table B – Debug Screen 5 OLC Values
Parameter
Qd
Dialysate Flow Rate
QdS
Dialysate Flow Stability
Current TI
Pre Dialyzer Temp sensor
NTC #3 stability
QbS
Blood Pump Stability Counter
Avg TI
Pre Dialyzer Temp stability
over the last 12 minutes
Pre Stbl Ct
Pre Dialyzer counter
Pre % Stbl
Percentage of time Pre
Dialyzer Cond was stable over
the last 12 minutes
Post Stbl Ct
Post Dialyzer Cond counter
Post % Stbl
Percentage of time Post
Dialyzer Cond was stable over
the last 12 minutes
OLC status
Current stage of OLC test
OLC Enable
OLC Enabled
CPRE Stable
Pre Dialyzer Conductivity
when stable during OLC test
Example: 15318 = 15.318 mS
CPOS Stble
Post Dialyzer Conductivity
when stable during OLC test
Example: 15318 = 15.318 mS
Description
2008T Troubleshooting Guide
P/N 490292 Rev. A
Should be within +/- 10 ml/min of the selected Dialysate
Flow rate
0 = Dialysate Flow is unstable
1 = Dialysate Flow is stable.
NOTE: QdS may = 0 if a Temp or Cond alarm occurs
OR if Dialysate Flow rate was changed OR if a Filling
Program occurs OR if a TMP alarm is reset
Ranges from 0 (unstable) to 100 (very stable)
Normally 0 or 1. If QbS reaches 20, the OLC test will
not start OR will cancel
When Dialysate Flow = 500 ml/min, should be more
than 70%. May be lower at other flow rates.
Defaults to 90. Decreases to 0 if Pre Dialyzer Cond is
stable
If OLC test is NOT running should be more than 90%.
Instability (i.e. less than 90%) is NORMAL within 12
minutes of an OLC test.
Defaults to 70. Decreases to 0 if Post Dialyzer Cond is
stable.
If OLC test is NOT running should be more than 90%.
Instability (less than 90%) is NORMAL within 12
minutes of an OLC test.
0 = An OLC test is NOT running
1 = At or going to approx.15.5 mS
2 = At or going to the approx. 13.5 mS
0 = OLC is turned off (OLC will NOT run).
1 = If OLC is turned on.
Not valid UNLESS an OLC test was performed.
Not valid UNLESS an OLC was performed.
Value
Page 434