Isolate Acid Pump Input - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

V105SC- 2.0.0 ISOLATE ACID PUMP INPUT
Reattach the Acid connector's inlet tubing!
a)
b) Remove the hydraulics from the cabinet.
c) Per the Figures below, THREE (3) checks:
CHECK #1: Check for restrictions between the acid (red) connector and Valve #105.
CHECK #2: ENSURE the
CHECK #3: ENSURE the Acid Pump's INPUT tubing is clear AND goes, unrestricted, to the bottom of
Pressure Sensor #106.
d) See procedure number
Acid Pump's
INPUT (white) nozzle is on the bottom.
V105SC- 2.1.0
(page 750).
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 749