DIASAFE- 5.0.0 ATMP STAYS 300 OR MORE / ISOLATE DIASAFE FILTER #92
a) Figure below, remove FILTER #92 from Valve #28's tubing.
Reset ALL alarms.
b)
c) Press the [Test & Options] tab.
d) Press the 'Diasafe Test' button then 'Enter' then allow the test to finish! TWO (2) possible scenarios:
1)
IF (and ONLY if) "Diasafe Test Passed": FILTER #92 MAY be restricted! Repeat the test
once more as the failure may be intermittent!
2) IF "Diasafe Test Failed": See procedure number
®
DiaSafe
Filter / Valve #28 / Filter #92
2008T Troubleshooting Guide
P/N 490292 Rev. A
DIASAFE- 6.0.0
(page 633).