Continue To Isolate Leak - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

LEAKING- 3.2.0 CONTINUE TO ISOLATE LEAK
With the machine OFF! FIVE (5) possible scenarios 1) or 2) or 3) or 4) or 5) below:
1)
Figure right, if (and ONLY if) the DiaSafe
headers: A) ENSURE the packing tabs are not present; B) Replace the filter!
2)
Per the Figure below, IF (and ONLY if) Flow Relief Valve #78 was leaking: See
procedure number
3)
IF (and ONLY if) the leak was from inside the DiaSafe
Inside the housing, tie wrap ALL tubing connections THEN see procedure
number
LEAKING- 4.0.0
4)
Per the Figure below, IF (and ONLY if) the leak was between the Flow Pump
and BALANCING CHAMBER VALVES #36 and/or #38: See procedure
number
LEAKING- 3.3.0
All OTHER scenarios:
5)
LEAKING- 3.3.0
(page 606)
(page 607).
(page 606).
See procedure number
2008T Troubleshooting Guide
P/N 490292 Rev. A
®
filter was leaking from its blue
®
filter housing:
LEAKING- 4.0.0
(page 607).