Step # 2: Read before parts a THROUGH e BEFORE performing them:
a) Turn the machine OFF then back on.
b) Return to Dialysis Program with [Dialysate Flow] set to 500 ml/min.
c) Allow six (6) minutes ENSURING the Home screen's [Conductivity] stabilizes
between 13.0 and 14.3 mS.
Reset all alarms!
d)
e) Read before performing! Return to
(and ONLY if) you return here continue to step #3.
Step #3: Read parts a THROUGH e below BEFORE performing them!
a) Perform an Acid Clean per procedure.
b) Return to Dialysis Program with [Dialysate Flow] set to 500 ml/min.
c) Allow six (6) minutes, ENSURING the Home screen's [Conductivity] stabilizes between
13.0 and 14.3 mS.
Reset all alarms!
d)
e) Read before performing! Return to
ONLY if) you return here proceed to
calibrate COND CELLS. After the calibration if (and ONLY if) you return here, THREE
(3) possible bad components listed below:
Component List: 1) Power Logic Board; 2) Actuator-Test Board; 3) Functional Board.
page 423
page 423
page 384
LEFT BLANK INTENTIONALLY
2008T Troubleshooting Guide
P/N 490292 Rev. A
, procedure number
CO-
, procedure number
CO-
, procedure number
CO- 6.2.0
10.0.40. If
10.0.40. If (and