Procedure # Co - 1.0.3 Continued; Equipped With Bibag Connector - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Procedure # CO- 1.0.3 continued:
H) FIRMLY connect BOTH concentrate connectors to
Return to Dialysis Program ("Select Program" → 'Dialysis' → 'Enter')
I)
J) From the Home screen, set [Dialysate Flow] to 800 ml/min and press 'Enter'.
K) Figure right, equipped with a bibag Connector?
Yes
See procedure number
(page 340).
No
NOT equipped with a bibag Connector!
Proceed to
CO-
1.0.3.7.
CO- 1.0.3.1 EQUIPPED WITH BIBAG CONNECTOR
A)
Call debug screen 15. Locate:
• V105 Er C (1
st
• V105 Er (2
nd
column from left). Should = 0 always!
• V104 Er C (3
rd
• V104 Er (3
rd
column from left). Should = 0 always!
B)
WITHOUT LOOKING AWAY, watch ALL FOUR, for four (4) minutes. If a '1' appears, even just once,
indicates a problem. Do ALL FOUR (4) REMAIN = 0?
Yes ALL FOUR ALWAYS = 0!
No
One or more = 1! FOUR (4) possible scenarios 1) or 2) or 3) or 4) below:
1)
IF (and ONLY if) V105 Er C EVER = 1 = Valve #105 possibly stuck closed: Proceed
page 435
to
2) IF (and ONLY if) V105 Er EVER = 1: Proceed to
OR VALVE 105
3)
IF (and ONLY if) V104 Er C EVER = 1 = Valve #104 possibly stuck closed: Proceed
page 438
to
4) IF V104 Er EVER = 1: Proceed to
ERROR
CO- 1.0.3.1
page 350
, procedure number
column from left). Should = 0 always!
column from left). Should = 0 always!
From the Home screen, if [Conductivity] is more than 13.4 mS
do not continue. If [Conductivity] remains less than 13.4 see
part C next page.
, procedure number
ERROR.
, procedure number
2008T Troubleshooting Guide
P/N 490292 Rev. A
JUGS
.
.
CO-
13.0.00.
page 744
CO-
14.0.00.
page 744
,
SECTION 31– VALVE 104 OR VALVE 105
,
SECTION 31– VALVE 104
Page 340