T- 1.4.0 [TEMPERATURE] NOT LESS THAN 30º C / ISOLATE DEBUG TEMP
With NTC #3 REMAINING unplugged, call debug screen 4. Is TEMP (lower left) less than
0.3?
Yes
TEMP less than 0.3! See procedure number
No
TEMP more than 0.3! Read before performing: A) Turn the machine off; B) Swap in the listed
components (see COMPONENT LIST below) one at a time, with known good, then in between;
C) Return to Dialysis Program and repeat procedure number
TEMP is less than 0.3 indicating the last component swapped in is the problem.
COMPONENT LIST: 1) Actuator-Test Board
4) Functional Board (possibly IC20)
1
To LOCATE these boards refer to
2
To prevent "Cond Offset Failure" place the machine into T and C Mode (refer to
MODES
(page 19))
T- 1.4.1 CHECK 'TEMP COMP' OPTION
Enter Service Mode → Options → Hardware Options. Does the "Yes" box
next to 'Temp Comp' have an 'X' in it (Yes or No)?
Yes
Place the 'X' in the "No" box and press 'Enter'. The 'X' turns blue.
Post a note that 'Temp Comp' is off then see procedure number
T- 1.4.2
(page 262).
No
Temp Comp = No! See procedure number
T- 1.4.2 TEMP COMP = NO / ATTEMPT TO SET TEMP LOW RANGE / TEMP CONTROL
NOTE!
This procedure is NOT a calibration! There is NO NEED to install a meter yet! DO NOT
touch the screen's [Temp DAC] data box!
a) With NTC #3 remaining unplugged, select Calibrate Sensors → Temp Control.
b) Press and release 'Enter' until the screen's [Monitor Reference] box turns pale yellow / white.
c) Press [Monitor Reference], it turns bright yellow.
d) Adjust [Monitor Reference] to as low as it will go then increase it 0.2° C ABOVE that.
e) Press and release the 'Enter' key until the calibration is saved.
f)
Turn the machine OFF!
g) Turn the machine on and return to Dialysis Program ("Select Program" → 'Dialysis' → 'Enter')!
h) With NTC #3 remaining unplugged, call the Home screen. Is [Temperature] less than 30.0° C now?
T- 1.4.1
1
; 2) Sensor Board
1,2
; 6) Distribution board; 7) Motherboard
Figure 4A
(page 10)
T- 1.4.2
(page 262).
2008T Troubleshooting Guide
P/N 490292 Rev. A
(page 262).
T- 1.4.0
(page 262) until screen 4's
1, 2
; 3) Sensor Board cable;
1
.