CO- 10.0.30 PROBLEM WITH POST DIALYZER COND
ENSURING the shunt door is open, is screen 0's Post Dialyzer Cond more than 10.000 mS?
Yes
Post Cond more than 10.000! See procedure number
No
Post Cond less than 10.000! See procedure number
CO- 10.0.32 POST DIALYZER COND MORE THAN 10.000
Is screen 0's Post Dialyzer Cond more than 20.000 mS?
Yes
Post Cond more than 20.000! See procedure number
No
Post Cond less than 20.000! Proceed to
CO- 10.0.34 ISOLATE POST COND CIRCUIT
a) Figure right, from the distribution board, unplug the
th
10
connector cap from the LEFT. This is Post
Dialyzer Cond Cell #13, "x13, COND-POS"!
b) Using a flashlight, check inside the "x13" position
for damage or corrosion.
c) Call debug screen 5. Is FPOS more
than 4000 (four thousand)?
Yes
FPOS more than 4000! Either Post Dialyzer Cond Cell #13 was not unplugged from
position "x13" or there is a problem with its circuit. THREE (3) possible bad
components': 1) Sensor Board
1
To prevent "Cond Offset Failure" place the machine into T and C mode. Refer to
OPERATING
No
FPOS less than 4000! See parts a AND b below:
a) Per the Figure right, place the
TWO-RESISTOR SET
vacant distribution board position, "x13, COND-POS".
b) Is FPOS between 13500 and 14500?
Yes
No
1
OR; 2) Sensor Board cable OR; 3) distribution board.
MODES, (page 19).
274 Ω
, into the Post Dialyzer Cond Cell's #13
FPOS between 13500 and 14500! See procedure number
(page 422).
274 Ω
Either the
resistor plug was not placed properly or there is a problem
with the Post Dialyzer Cond Circuit: THREE (3) possible bad components:
a
1) Sensor Board
OR 2) Sensor Board cable OR; 3) Distribution board.
a
To prevent "Cond Offset Failure" place the machine into T and C mode.
Refer to
OPERATING MODES
2008T Troubleshooting Guide
P/N 490292 Rev. A
CO- 10.0.32
CO- 10.0.34
CO- 10.0.34
page 384
, procedure number
resistor plug, from the
(page 19).
(page 421),
(page 421),
(page 421).
CO-
6.2.0.
CO- 10.0.36
Page 421