V104SC- 1.0.22 BIC PRESS NOT BETWEEN -50 AND 30
A) Leaving the machine on, pull the hydraulics away from the cabinet to.
B) Figures below, per the numeric tubing identification (ID) bands, ENSURE the bibag Connector tubing
segments are not reversed or kinked!
C) Is Bic Press now between negative (-)50 and positive 30 (Yes or No)?
Yes
Bic Press between negative 50 and positive 30! The tubing may have been restricted:
A) Reinstall the hydraulics ENSURING the tubing remain unrestricted! B) Return to
(ABOVE) procedure number
No Bic Press still is NOT between negative 50 and positive 30! See part D next page.
V104SC- 1.0.0
(page 730).
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 732