Fresenius Medical Care 2008T Troubleshooting Manual page 739

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

A)
Press 'Esc' to return to the main RINSE screen. THREE (3) possible scenarios below:
1) IF (and ONLY if) the [Remaining Time] window is NOT 'blinking': See parts B)
and C).
2) IF (and ONLY if) [Remaining Time] is 'blinking' AND the "V104/108 Stuck
Closed" banner is up: Proceed to
V104SC-
2.0.3.
3) IF [Remaining Time] is 'blinking' AND the "V104/108 Stuck Closed" banner is
NOT up: Address whatever alarm banner is showing.
B)
The first part of clearing "V104/108 Stuck Closed", is successful. Now when rinse finishes,
Bic Press must significantly decrease through Valves #104 and #108.
C)
Watch carefully until if [Remaining Time] = 0:00. Does "V104/108 Stuck Closed" reappear?
Yes
"V104/108 Stuck Closed" reappears! Proceed to
number
No
"V104/108 Stuck Closed" NEVER reappears! When RINSE finishes return
to Dialysis Program ("Select Program" →.'Dialysis' → 'Enter'). TWO (2)
possible scenarios:
1)
2) IF V104/108 Stuck Closed" does NOT reoccur: The alarm has been
cleared from memory! Possibly the blue connector was not inserted
properly into its rinse port when last rinsed.
page 739
V104SC- 2.0.3.
IF (and ONLY if) "V104/108 Stuck Closed" reoccurs: Proceed to
page 438
, procedure number
NO TEXT INTENIONALLY
2008T Troubleshooting Guide
P/N 490292 Rev. A
, procedure number
page 739
, procedure
CO-
14.0.0.