Less Than 180 Ml Collected - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

V104SC- 2.0.6 LESS THAN 180 ML COLLECTED
A) Return the blue connector the Bicarbonate
Rinse port.
B) Figure right, REMOVE the
from the hydraulic side of the
RINSE PORT
and direct it into the
cylinder.
C) Expecting flow from the tubing,
RINSE!
D) If flow, allow three (3) pulses THEN measure
ten (10) pulses.
E) Interrupt Rinse by pulling the blue connector
out of the Rinse port!
F) More than 180 ml collected?
Yes
More than 180 ml! TWO (2) possible bad components: 1) Restricted Bicarb concentrate
connector OR; 2) Bad Bicarb Rinse port.
No
Less than 180 ml! ENSURING the machine was in RINSE prior to checking flow, perform parts
a AND b below:
a) With RINSE now interrupted, reattach the Bicarb rinse port tubing!
b) FIVE (5) possible problems: 1) Figure below, bad wire connection(s) between Valve #108
and Valve #104 OR; 2) Loose ribbon cable at the bibag Interface board (refer to
(page 11)) OR; 3) Bad Valve #108 OR; 4) Bad bibag Interface Board OR; 5) Bad electrical
connection, from Valve #108/Valve #104, inside Distribution Box #2 (see
13)).
WHITE
tubing
BICARB
return to
2008T Troubleshooting Guide
P/N 490292 Rev. A
Figure 4C
Figure 4D