Isolate Dialysate Settings; Dialysate Screen View - Fresenius Medical Care 2008T Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

CO- 1.0.0 ISOLATE DIALYSATE SETTINGS
A) At the bottom of the screen, press the 'Dialysate' tab.
B) Figure below, ENSURE the 'Selected Acid' matches the acid connected to the machine.
C) Set [Base Na+] to 137 and press 'Enter'.
D) Set [Bicarbonate] to 33 and press 'Enter'.
E) From the Home screen, ENSURE [Dialysate Flow] is on to at 500 ml/min or more.
F) If the 'Selected Acid' AND / OR [Base Na+] AND / OR [Bicarbonate] was changed, allow five (5)
minutes BEFORE continuing to part G.
G) Based on [Conductivity] now, THREE (3) possible scenarios:
1)
IF (and ONLY if) LESS THAN 13.0 mS: See procedure number
2)
IF (and ONLY if) STABLE between 13.0 and 14.3: Proceed to
CO-
1.0.9.
3)
ALL OTHERS: Proceed to
Figure 55 – Dialysate Screen
page 353
, procedure number
2008T Troubleshooting Guide
P/N 490292 Rev. A
CO- 1.0.1
(page 338).
page 363
, procedure number
CO-
1.0.4.
Page 337