CLEAN- 1.2.6 MOTOR CONTINUES TO ROTATE
Recheck Loading Pressure (gauge in Rinse port). THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) ALWAYS less than 15 psi: Leaving the machine in the Cleaning Program, see
procedure number
2) IF (and ONLY if) MORE than 25 psi: Figure below, turn Valve #65's nut counterclockwise (outward) to
adjust to between 23 and 25 psi then proceed
3)
IF cycling to between 15 and 21 psi: Figure below, turn Valve #65's nut clockwise (inward)
attempting to adjust pressure to a PEAK between 23 and 25 psi. Can you?
Yes
Pressure CAN be adjusted to its PEAK! Allow two (2) minutes then if a Flow Error
reoccurs proceed
No
Loading Pressure remains low! Leaving the machine in the Cleaning Program, see
procedure number
CLEAN- 1.5.0 LOADING PRESSURE REMAINS LOW
a) Turn the Heater Breaker Switch OFF!
b) Figure right, unplug the Deaeration Pump from
distribution board position, "P20, DEGAS-P".
1000 ml
c) Obtain a
the white drip bucket.
d) See procedure number
CLEAN- 1.5.0
(page 176).
page 190
, procedure number
CLEAN- 1.5.0
graduated cylinder AND
CLEAN- 1.5.1
(page 177).
2008T Troubleshooting Guide
P/N 490292 Rev. A
page 190
, procedure number
CLEAN-
(page 176).
CLEAN-
3.0.0.
3.0.0.
Page 176