DIASAFE- 7.0.0 VALVE #28 OKAY
a) Return FILTER #92 to Valve #28!
Per the Figure below
b)
#26!
c) Are you ABSOLUTELY SURE the DiaSafe
Yes
Tubing is plumbed properly! See procedure number
No Tubing is backwards! Per the Figure, attach the tubing properly and repeat the test.
, ENSURE 'TUBE A' goes to VALVE #24 AND 'TUBE B' goes to VALVE
®
tubing is plumbed correctly?
2008T Troubleshooting Guide
P/N 490292 Rev. A
DIASAFE- 8.0.0
(page 636).